SIGNAL
PROCESSING
A
low cost signal processing
scheme has been developed. The method of driving and receiving used
in the flowmeter employs reciprocal action. This is essentially
driving the transducers from and receiving the signal into the same
impedance. In this case we are driving from, and receiving into, a
low impedance. This approach ensures that the transducers at zero
flow have identical receive signals independent of matching of the
transducers, coupling and liquid temperature. The only requirement
for this to work is that the transducers and associated circuitry
should be linear and that during reception the transducers do not see
any signal generated by themselves since this signal is not
reciprocal.
The
transducers are thus driven from the output of a low impedance linear
amplifier and the signal received from the transducers is received
into a low impedance transimpedance amplifier. The signal is then
passed through a band pass filter centred at the drive frequency
before being input to a 2MHz 12bit ADC. Electronic switches in the
system change the roles of the transducers from transmitter to
receiver and the upstream and downstream transducer are excited
sequentially. The data from the ADC in both directions is averaged
over a number of transmissions and stored.
A
novel time measurement algorithm then processes the signals to
extract the time difference between them. This signal is then
processed to compensate the measurement for the speed of sound in the
fluid which enables the flowmeter to be used with a range of fluids
and also to be independent of the speed of sound in the fluid. The
electronics is capable of providing the user with a visual indication
of flow, a pulse output giving a pulse per unit volume of flow or a 4-20mA output.
PERFORMANCE
Figure
8 shows a typical performance of a 6mm flowmeter. The calibration
shows that the flowmeter is capable of measuring from 1.5m3/hr to approximately 3l/hr. Over a range of 100: 1 it is well within +/-2%.
By
changing the bore of the meter to 1mm flows as low as 0.2l/hr have
been measured.

Calibration Characteristics of Flowmeter
|